Q: What about NIST cal on TOPAZV and ETS7000?
NIST Calibration on Legacy Systems
Equipment
Each instrument should be warmed up for an hour prior
to the calibration using the same AC outlets that will be used during the
calibration in order to prevent destabilization by unplugging and moving
the instrument.
Click your browser's Back button to return to the Q'nApps index. |
NIST Calibration Equipment List It is important to note the DUT board used. Do not use adapter type DUT boards as the power supply sense lines are not connected to the DUT power on these boards. Most labs have cooling systems, often with floor vents. If your air conditioning turns on and off you should position the test system and instruments so that they are out of drafts to insure stability during the procedure. HiLevel recommends NIST calibration to be performed every six months. If you have further questions or comments about the calibration, email us or call 1-800-HILEVEL. QL40.zip is a zipped Word file of this Q'nApp. |